Non warranty replacement regulator for the Apyx Renuvion device. Please verify with your local Apyx representative prior to purchase.
INDICATIONS FOR USE:
The HELIUM GAS REGULATOR is for use with the electrosurgical generators. The regulator is used to provide a constant flow of helium at approximately 36 psi / 248 kPa to the generator. The helium gas is stored in a gas tank vessel under high pressure. The Helium Gas Regulator reduces the high pressure (about 500-4000 psi / 3447-27579 kPa) to a lower pressure (about 36 psi / 248 kPa). The regulator has an internal safety relief valve to prevent over-pressurization of the regulator.